632.51
WIKA 632.51 – (Capsule Pressure Gauge) WIKA/Germany
Applications
- Pressure measurement at very low pressures
- For gaseous, aggressive media, also in aggressive environments
- Robust design and ingress protection IP54, suitable for outdoor use
Special Features
- High overload safety up to 50 x full scale value
- Scale ranges from 0 … 2.5 mbar
- Measuring chamber protected against unauthorised intervention
- Low measuring error and influence on function from medium pollution
Description
These capsule pressure gauges are used when extremely-low pressures must be displayed on site. The instruments can be overloaded up to 50 times the full scale value without any problems.
The large effective area of the two welded, radially corrugated diaphragms of the capsule element allows the reliable measurement of the smallest pressures.
These measuring instruments, made entirely of stainless steel, are suitable for gaseous, aggressive media.
The capsule pressure gauge of model 632.51 has been qualified and manufactured in accordance with EN 837-3.
Design | |
Nominal size in mm | 100, 160 |
Accuracy class | 1.6 |
Scale ranges | 0 … 2.5 to 0 … 100 mbar
or all other equivalent vacuum or combined pressure and vacuum ranges |
Pressure limitation | Steady: Full scale value
Fluctuating: 0.9 x full scale value |
Permissible temperature | Ambient: -20 … +60 °C
Medium: +100 °C maximum |
Temperature effect | When the temperature of the measuring system deviates from the reference temperature (+20 °C): max. ±0.6 %/10 K of full scale value |
Process connection | Stainless steel 1.4571,
lower mount, G ½ B (male), SW 22 |
Pressure element | Stainless steel 1.4571 |
Movement | Stainless steel |
Dial | Aluminium, white, black lettering |
Pointer | Adjustable pointer, aluminium, black |
Case | Stainless steel, with blow-out device |
Window | Laminated safety glass |
Options | ■ Other process connection
■ Sealings, see data sheet AC 09.08 ■ Instrument mounting bracket for wall or pipe mounting, see data sheet AC 09.07 ■ Panel or surface mounting flange (consider measuring chamber!) ■ Indication accuracy class 0.6 or 1.0 1) ■ Higher overload safety 1) ■ Capsule pressure gauge with switch contacts, see data sheet PV 26.06 ■ Capsule pressure gauge with output signal, see data sheet PV 16.06 |