TR7439
Evaluation unit with display for PT100/PT1000 temperature sensors TR7439
Product characteristics | |||||||
---|---|---|---|---|---|---|---|
Number of inputs and outputs | Number of digital outputs: 2 | ||||||
Measuring range |
|
||||||
Process connection | threaded connection G 1/2 external thread | ||||||
Application | |||||||
Special feature | Gold-plated contacts | ||||||
Application | for Pt100 and Pt1000 measuring elements | ||||||
Electrical data | |||||||
Operating voltage [V] | 18…32 DC; (“supply class 2” to cULus) | ||||||
Current consumption [mA] | < 50 | ||||||
Protection class | III | ||||||
Reverse polarity protection | yes | ||||||
Power-on delay time [s] | 1 | ||||||
Integrated watchdog | yes | ||||||
Inputs / outputs | |||||||
Number of inputs and outputs | Number of digital outputs: 2 | ||||||
Outputs | |||||||
Total number of outputs | 2 | ||||||
Output signal | switching signal; IO-Link; (configurable) | ||||||
Electrical design | PNP/NPN | ||||||
Number of digital outputs | 2 | ||||||
Output function | normally open / normally closed; (parameterisable) | ||||||
Max. voltage drop switching output DC [V] | 2.5 | ||||||
Permanent current rating of switching output DC [mA] | 250 | ||||||
Short-circuit protection | yes | ||||||
Type of short-circuit protection | pulsed | ||||||
Overload protection | yes | ||||||
Measuring/setting range | |||||||
Measuring range |
|
||||||
Set point SP |
|
||||||
Reset point rP |
|
||||||
In steps of |
|
||||||
Resolution | |||||||
Resolution of switching output [K] | 0.1 | ||||||
Resolution of display [K] | 0.1 | ||||||
Accuracy / deviations | |||||||
Switch point accuracy [K] | ± 0,3 + (± 0,1 % MS) | ||||||
Display accuracy [K] | ± 0,3 + (± 0,1 % MS) | ||||||
Temperature coefficient [% of the span / 10 K] | 0,1; (in case of deviation from the reference condition 25 ± 5 °C) | ||||||
Response times | |||||||
Response time [ms] | 390 | ||||||
Software / programming | |||||||
Parameter setting options | hysteresis / window; normally open / normally closed; switching logic; switch-on/switch-off delay; Damping; Display unit | ||||||
Interfaces | |||||||
Communication interface | IO-Link | ||||||
Transmission type | COM2 (38,4 kBaud) | ||||||
IO-Link revision | 1.1 | ||||||
Operating conditions |
|||||||
Ambient temperature [°C] | -25…80 | ||||||
Storage temperature [°C] | -40…100 | ||||||
Protection | IP 67 | ||||||
Tests / approvals | |||||||
EMC |
|
||||||
Shock resistance |
|
||||||
Vibration resistance |
|
||||||
MTTF [years] | 252 | ||||||
UL approval |
|
||||||
Mechanical data | |||||||
Weight [g] | 198.5 | ||||||
Materials | stainless steel (1.4404 / 316L); stainless steel (1.4301 / 304); PBT+PC-GF30; PBT-GF20; PC | ||||||
Process connection | threaded connection G 1/2 external thread | ||||||
Displays / operating elements | |||||||
Display |
|
||||||
Remarks | |||||||
Remarks |
|
||||||
Pack quantity | 1 pcs. | ||||||
Electrical connection | |||||||
Connection | Connector: 1 x M12; Contacts: gold-plated |